Archive for March 2016

B&K probe checking for leaks

Some B&K probes should be fully water proof even on the connector (with connector sealing plate installed). You should be pressurizing the probe and checking for leaks in a water bath. This is what it will look like if you find a leak!

Read More

Backup your ultrasound machine

Are you a small ultrasound practice and you don’t have a digital repository for your images? Ultrasound systems are not designed to be used as digital archiving devices. Remember to backup your patient studies. Alternately, talk to us today about affordable ways you can download your images to a PC for safe reliable storage.

Read More

Acuson/Siemens V5Ms Cardiac Phased array probe image degradation

Currently we have a probe in our laboratory for investigation. It is a Siemens/Acuson V5Ms TOE/TEE probe in which the customer was complaining about poor image quality. Our acoustic report suggests the probe has approximately 10% of elements non firing as can be seen in the image below   The issue is almost certainly caused […]

Read More

Air under the lens of a TEE/TOE probe

This video shows the air under the lens of a particularly grubby Philips S7-2t transesophageal echocardiogram (TOE/TEE) probe as it rotates. The air results in poor image quality. If you have one of these, check it for air and let us know if you have any problems at www.probelogic.com.au

Read More

Before and after recoating and relabeling of a TEE insertion tube

We recieved a GE 6VT 3d TOE/TEE probe in for repair and it had severe insertion tube damage as can be seen below. After the ProbeLogic Recoat/Relabel process the probe is back to new condition Check your probes and let us know if we can assist you in bringing your probe back to new condition […]

Read More

Frightening! – TEE probe with fine wire protruding

Today’s post is a frightening one. We received a TEE/TOE probe in for repair. Protruding from the side of the bending rubber section was a very fine piece of stainless steel wire which had broken away from the articulation wires. The thought of this being inserted into somebody’s esophagus with this protruding from the side […]

Read More