What is a matching layer?

You have probably heard us speak of a matching layer between the acoustic array and the lens of the probe. What is this mysterious layer? The matching layer helps to couple the acoustic impedance of the lens material with the acoustic impedance of the array itself. This avoids unnecessary reflections at this mismatched impedence boundary. […]

Read More

What happens when water gets in the TIP of a TEE/TOE probe?

When water enters the tip of a TOE/TEE probe, usually through a hole in the bending rubber it causes short circuits between elements. This causes several strange artifacts and image degradation. It can clearly be seen as an increase in capacitance on the electro-acoustic test. ProbeLogic can restore a probe suffering from these faults to […]

Read More

More fluid under the lens

Here is another example of fluid under the lens of a probe. This time a GE M3S probe. If you have any doubt as to whether there is a ‘leak’ in your lens. Let us know!  