Probe case seperating? Check out whats inside!

We currently have a probe in the lab for poor imaging. On initial inspection all seemed fairly normal but with a slight separation of the seams of the probe.

IMG_2689

After removing the covers look at the amount of ultrasound gel and other fluid that has been forced inside the probe.

IMG_2852

IMG_2861

This amount of fluid inside the probe for such a long period of time resulted in all of this corrosion.

IMG00196 IMG_2894 IMG_2895 IMG_2896

This damage can often be repaired but is quite costly. Avoiding this cost is as simple as sending your probe to us as soon as you notice the gap in the cases. Inspect your probes today.